8. Troubleshooting

Use the table below to diagnose and resolve common field issues. If the issue cannot be resolved, contact VENTMEDiC® customer support and do not open the enclosure or service internal components without authorization.

SymptomPossible CauseCorrective Action
LED buttons do not blink on power-upPower switch not fully engaged; depleted batteryRe-press power switch firmly; check battery charge via Web App; connect charger if needed
Device shows Offline in Web AppAntenna not deployed; no cellular coverage; Wireless Lockout Key Switch is OFFVerify antenna is upright and connected; confirm key switch is ON; connect Ethernet/Starlink if no cellular
No improvement in flow readings after switching to FLOW modeValve still actuating; gas source not flowing; loose gas fittingWait for LED to stop flashing and confirm mode is active; verify vent source is in stabilized flowing state; inspect all gas fittings
Pressure builds unexpectedly in VENT modeUpstream valve partially open; outlet partially blockedCheck upstream isolation valve; verify outlet hose is unobstructed and properly routed
Battery not charging from solar panelPanel not correctly oriented; panel surface dirty or shadedAdjust panel to ~45° south-facing; clean panel surface; remove obstructions
Enclosure door will not seal properlyDebris on gasket; gasket damagedClean gasket surface; inspect for cracks or deformation; contact VENTMEDiC® if gasket is compromised
Abnormal odour or suspected gas leakLoose fitting; damaged hose or sealImmediately stop measurement, move to a safe location, and follow site emergency procedures; do not re-enter until cleared by site authority
⚠️

IMPORTANT: If the issue cannot be resolved using this table, contact VENTMEDiC® customer support. Do not open the enclosure or service internal components without authorization.